8 - Maintaining the instrument

219

 Lift off the ring at the pilots.

 Clean the plastic top section and ring with

a soft cloth or bottle brush and

some water with a mild laboratory cleaner.

 Allow the carousel to dry. If necessary, remove the last traces of moisture with warm air.

 Replace the ring on the plastic top section and use a wrench to turn the hexagon head screws so

that the tips of the screws prevent you from lifting off the ring (tighten the screws only slightly).
The metal ring will be held with the plastic top section by means of a magnet.

 Move the plastic section until the reference hole (bore in the plastic section without number

indicated by a white ring) matches the bore in the metal ring.

 Insert the carousel into the holder and observe that the ejector openings of the ring must be

above the ball valve.

 Subsequently, perform a reference run.