Specify lower and upper calibration range (dialog), H_742, O_4123 – BUCHI DuMaster D-480 User Manual

Page 337: H_740, H_806, H_821

Advertising
background image

Operation Manual D-480

©BÜCH Labortechnik AG

10 - Menu and dialog descriptions

337

Further information

Further information includes instructions in which the dialog is used and knowledge articles
concerning the dialog:

Defining standard substances as calibration samples (on page 122).

Background knowledge required for calibration (on page 52).

Calibration formulae (on page 55).

Specify lower and upper calibration range (Dialog)

Function

In this dialog you specify for a two-range calibration at what point to divide between the upper and
lower calibration range. The dialog displays all samples on the basis of a virtual line. This shows you
whether it makes sense to split the ranges with a dividing line.

Open

Open the dialog via Calibration > Calibrate. First the "Configure calibration" dialog is displayed in
which you specify the calculation methods. Only then the "Specify lower and upper calibration range"
dialog is displayed if you have specified a two-range calibration.

Elements

The following describes the elements of the dialog.

Label

Element type

Meaning

<<<

Button

Moves the dividing line between the upper and lower
calibration range to the left in larger intervals.

<

Button

Moves the dividing line between the upper and lower
calibration range to the left in smaller intervals.

>

Button

Moves the dividing line between the upper and lower
calibration range to the right in smaller intervals.

>>>

Button

Moves the dividing line between the upper and lower
calibration range to the right in larger intervals.

Zoom +

Button

Increases the scale of the view.

Zoom -

Button

Decreases the scale of the view.

Zoom

Label

Displays the current zoom setting.

Dividing line at

Label

Displays in figures where the dividing line is currently set
between the upper and lower calibration range.

OK

Button

Saves new settings

Cancel

Button

Aborts the process.

Note: The dividing line can also be set directly: To do so, place the mouse cursor at the point where
you want to place the dividing line and then click the right mouse button.

Further information

Further information includes instructions in which the dialog is used and knowledge articles
concerning the dialog:

Performing calibration.

Background knowledge required for calibration (on page 52).

Right mouse button function (on page 69).

Calibration formulae (on page 55)