background image

Operation Manual D-480

©BÜCH Labortechnik AG

10 - Menu and dialog descriptions

292

Label

Element type

Meaning

Cancel

Button

Aborts the process.

Further information

Further information includes instructions in which the dialog is used and knowledge articles
concerning the dialog:

Stopping continuous analysis (on page 250).

Right mouse button function (on page 69).

Adjust carousel position (Dialog)

Note

This function is only available in the solides modes.

Function

After a power failure or if you repeat initial startup and remove the carousel, it is necessary to readjust
the position of the carousel.

In the dialog:

you are warned about adjusting the carousel position if the carousel contains samples,

you can move the carousel to zero position or any other position.

Please observe the following instruction:

Caution

Inserting multiple samples into the ball valve (only for

solid operation)

If there are samples in the carousel while you are changing the position of the
carousel via System > Carousel position, multiple samples may fall into the ball
valve at the same time. In this case, a service technician is required.

Before changing the position of the carousel, always remove all samples from the
carousel.

Note

Observe the following when adjusting the position of the carousel:

The carousel is moved so that the selected position is above the insertion opening. When the next
analysis starts, the carousel moves the next sample position above the insertion opening and drops
the sample. Therefore, it makes sense to select the previous position when moving the carousel.

Example: You want to move the carousel so as to analyze the sample at position 26 in the next
analysis. Therefore, select position 25. The carousel moves so far that position 25 is above the
insertion opening. When the next analysis starts, the carousel moves one position on during the
insertion process and thus drops the samples at position 26.

Open

Open the dialog via System > Sample Position.

Elements

The following describes the elements of the dialog.

Label

Element type

Meaning

All samples removed
from the carousel

Checkbox

If you have removed all samples from the carousel, you
must check this box. Otherwise the carousel will not move.