Operation Manual D-480

©BÜCH Labortechnik AG

8 - Maintaining the instrument

190

 Press the post combustion tube plug (1) and the combustion tube connection (3) down lightly and

pull the furnace out of the instrument to the front.

 Pull the combustion tube connection (3) out of the combustion tube (4) with careful jiggling

motions.

 Place the combustion tube connection (3) on the tube rack.

 Insert the tube extractor into the combustion tube (4). After clamping the crucible tongs in the ash

crucible, pull the ash crucible out.