O_460, O_461, O_462 – BUCHI DuMaster D-480 User Manual

Page 54

Advertising
background image

Operation Manual D-480

©BÜCH Labortechnik AG

4 - Understanding the instrument and planning its use

54

Linear calculation over the whole range

The following picture shows an example in which a linear calculation over the whole range achieves
acceptable calibration curve quality.

Linear calculation over two ranges

The following picture shows an example in which a linear calculation over two ranges achieves
acceptable calibration curve quality.

Linear and polynomial calculation

The following picture shows an example in which the calibration curve exhibits acceptable quality by
means of a linear calculation in the low concentration range and a polynomial calculation in the higher
concentration range.