Method (dialog), H_1955, H_875 – BUCHI DuMaster D-480 User Manual
Page 315: H_1957, H_561, O_3769
Operation Manual D-480
©BÜCH Labortechnik AG
10 - Menu and dialog descriptions
315
Label
Element type
Meaning
Peak anticipation
Textbox
Field to input a time. Defines the time the system waits
for the occurrence of the nitrogen peak. If the system has
not detected a peak start after expiration of this peak
expectation time, it assumes that the sample does not
contain nitrogen.
Integrator
reset delay
Textbox
Field to input a time. Defines the time after which the
integrator is reset to zero before beginning actual
integration.
Max. integration time
Textbox
Field to input a time. Defines the maximum duration of
the peak integration.
Remaining elements:
Label
Element type
Meaning
OK
Button
Saves new settings
Cancel
Button
Discards settings not yet saved.
Help
Button
Opens the description of the dialog in the help.
Further information
Further information includes instructions in which the dialog is used and knowledge articles concerning the dialog:
Heating up the furnace / checking parameters (on page 118).
Replacing fuses (on page 243).
Shutting the instrument down for long measuring breaks (switching off) (on page 117).
Setting device parameters (on page 50).
Optimizing basic instrument settings (on page 99).
Method (Dialog)
Function
In this dialog you can create, delete, view, copy and change methods.
Note
The method parameters can not only be defined via a method but also via the "Device parameters"
dialog. If not otherwise defined by a method, the settings in the "Device parameters" dialog apply. If a
parameter is defined differently in the method and in the "Device parameters" dialog, the method
takes priority if it is used for the measurement.
Open
Open the dialog by selecting Options > Settings > Methods.
Elements
The following describes the elements of the dialog.
Label
Element type
Meaning
---
Listbox
This field lists all methods already defined.
Identifier
Textbox
Field to input the name of the method. When you enter the
name in the sample view, this method is allocated to the
appropriate sample.
Description
Textbox
Field to input the purpose of the method. For easier
selection of a method.