Set stop tag (dialog), Select current sample (dialog), O_4105 – BUCHI DuMaster D-480 User Manual

Page 278: O_4107

Advertising
background image

Operation Manual D-480

©BÜCH Labortechnik AG

10 - Menu and dialog descriptions

278

If 21 CFR Part 11 functionality is enabled, a new version of the document is created the next time it is
saved.

Set stop tag (Dialog)

Function

In this dialog you can specify the sample at which to set the stop marker. Usually, the program
automatically moves the stop marker while you input sample data. Therefore it is only necessary to
move the stop marker manually in special cases.

Open

Open the dialog:

via the menu command Edit > Stop tag or

by clicking the right mouse key to the corresponding sample cell. Select Set stop tag... from the
popping-up context menu.

Elements

The following describes the elements of the dialog.

Label

Element type

Meaning

Stop marker at:

Textbox with
buttons

Field to input the sample number at which to set the stop
marker.

You can also set the sample number in the textbox using
the arrows.

OK

Button

Saves the setting.

Cancel

Button

Aborts the process.

Further information

Further information includes instructions in which the dialog is used and knowledge articles
concerning the dialog:

Stopping continuous analysis (on page 250).

Right mouse button function (on page 69).

Select current sample (Dialog)

Function

You set the number of the current sample in the dialog. The current sample is the sample that is
currently being analyzed (analysis mode) or that will be analyzed next (idle mode). It is only
necessary to set the current sample after breakdowns (power failure, etc.), for example if the carousel
position no longer matches the serial sample number.

Note

The current sample is only set in sample memory. Therefore, it is necessary to check that the data of
the current sample in sample memory match the data of the sample on the carousel that will actually
be analyzed next. Otherwise, measurements will be incorrect.

Open

Open the dialog:

via the menu command Edit > Current sample or

Advertising