Calibration table for nitrogen calibration

Suggestion

The following table is a recommendation for a calibration. The user can customize settings to the
specific analytical task.

Number Sample type

Substance/
Sample weight

Absolute
content

Name

Method

2

Conditioning

Aspartic acid /
300 mg ±10 mg

RunIn

CalibrationLow

1

Conditioning

Aspartic acid /
10 mg

RunIn

CalibrationLow

2

Conditioning

Aspartic acid /
1 mg

RunIn

CalibrationLow

20

Calibration
samples

Aspartic acid /
1, 2, 3, 4, 5, 7 ,10,
15, 20, 25, 30, 50,
75, 100, 125, 150,
175, 200, 250,
300 mg

1 mg

0.105 mg N

Aspartic acid

CalibrationLow

4

Calibration
samples

Aspartic acid /
350, 450, 550,
650 mg

650 mg

68.38 mg N

Aspartic acid

CalibrationHigh