H_824, O_827, H_826 – BUCHI DuMaster D-480 User Manual

Page 180

Advertising
background image

Operation Manual D-480

©BÜCH Labortechnik AG

8 - Maintaining the instrument

180

 Click OK to save your settings and close the dialog. The "Specify lower and upper calibration

range" dialog appears. It displays a graph generated from the absolute contents of the samples
and their peak areas in a linear calculation. The graph displays deviations of the measuring
points from these lines.

 Observe the measuring points and decide

what calculation would be suitable

whether you would separate the the ranges

and, if so, where you would separate the ranges.

See Calibration curve calculation method criteria (on page 53).

 Write down your considerations.

 Click OK to view the readings for the other elements. For each element, decide as described

before.

 Write down your considerations.

 Click Cancel.

Phase 3: Set calculation methods, perform calculation

Proceed as follows:

 Execute the following command again: Math. > Calibration. If sequential blank value

determination is activated a message window opens showing the corresponding information.

 Click Yes to continue. The "Configure calibration" dialog appears.

 Define the settings that you ascertained in the preceding phase.

 Click OK to save your settings and close the dialog.

 If you have decided to split the calibration into a lower and an upper range the "Specify lower and

upper calibration range" dialog appears. Set the dividing line by clicking the < and > buttons and
then click OK.

 First, the calibration curve for the lower range appears in the calibration view.

Phase 4: Optimizing calibration curves and apply calibration

Background information can be found in Assessing the calibration curves (on page 182).

In order to optimize the calibration curve for each element you must exclude samples that deviate
significantly from the curve. Deviations of up to ± 1.3 µg are tolerable.

Proceed as follows:

 Zoom in/out the graph by selecting View > Zoom in/out so that you can check the overall curve.

 Click the

and

buttons to move from sample to sample in the calibration curve and

check the deviation "Diff.". Samples that display excessive deviation must be deleted.

 Select a sample to delete. Then click the "Include/Exclude" button"

. The deleted sample is

highlighted red.

 Compare "r" (new value) and "r_old" (old value) of the calibration curve and check whether

deleting has had the desired effect. See Assessing the calibration curves (on page 182).