Operation Manual D-480

©BÜCH Labortechnik AG

8 - Maintaining the instrument

209

Installing the Ni flap

Note

The following section is a continuation of "Removing the Ni flap". All steps described in "Removing the
Ni flap" must already have been performed. See: Removing and cleaning the Ni flap (on page 206).

Process

The section is divided into the following phases:

Preparing the instrument, gas pressure

Installing the Ni flap

Follow-up instrument preparation, gas pressure / gas-tightness

Safety instruction

Strictly observe the safety instruction "Hot instrument parts" under Warning notes during operation
(on page 24).

Phase 1: Preparing the instrument, gas pressure

Safety instruction

Strictly observe the safety instruction "Gas pressure" under Warning notes during operation (on
page 24).

To prepare the instrument for maintenance work follow the instructions in "Preparing the instrument:
Cool down the furnace, gas pressure" in Preparing and following up maintenance work (on page
171).

Top view furnace

The following picture shows a top view of the furnace pulled out:

Legend:

1 Post combustion tube plug
2 Post combustion tube
3 Combustion tube connection
4 Combustion tube
5 Reduction tube plug
6 Reduction tube

Phase 2: Installing the Ni flap

Proceed as follows:

 Place the combustion tube connection (3) onto the combustion tube (4). The lance must extend

into the ash crucible approx. 5 - 10 mm.

Advertising