Operation Manual D-480

©BÜCH Labortechnik AG

6 - Starting up and shutting down the instrument

115

CO

2

intake pressure

CO

2

intake pressure is set at the delivery point so that the following values are displayed on screen:

The "Input" display shows approx. 1300 mbar.

The "MFC CO2" display shows approx. 650 ml/min.

The "Flow CO2" display shows approx. 650 ml/min.

Flow rate

The following flow rate settings are typical:

The "MFC CO2" display shows approx. 650 ml/min for the measuring gas outlet.

The values for the reference gas outlet are between 30ml/min and 60ml/min. There is no display
for the reference gas outlet.

During standby, the "MFC O2" display shows 0 ml/min. The corresponding value will be
displayed only during O

2

dosing (combustion).