Operation Manual D-480

©BÜCH Labortechnik AG

5 - Work performed by the system administrator

78

What can you modify in the configuration?

Safety instruction

Strictly observe the safety instruction "Improper changes" under Warning notes during operation
(on page 24).

What can you set?

You can set the following in the configuration:

Interface properties

Weighing data properties

Signature program properties

Device parameter

See Configure analyzer (on page 77).

Defining logon timeout

What is the logon timeout?

The logon timeout defines the time to wait in which the user does not perform any action before the
operating software displays a dialog to log in again.

Safety instruction

Strictly observe the safety instruction "Improper changes" under Warning notes during operation
(on page 24).

Defining logon timeout

Proceed as follows:

 Open the "Device configuration" dialog by selecting Options > Configuration.

 Select the "DEVICE.INI" tab. Select the "System" section by clicking the arrow next to the

"Section" dropdown listbox.

 Select in the "Items" listbox the entry "Logon Timeout" and click Edit. The "Configure key value"

dialog is displayed:

 Enter the number of seconds to wait before displaying the login dialog again in the "Value" field

and click OK. The "Comment modification" dialog is displayed.

 Enter a justification in the "Comment modification" dialog and click Close.

 Click OK in the "Device configuration" dialog. Your settings are saved and the dialog is closed. A

message window appears which informs that the modification will only become operative after
the instrument software has been left and re-started.

 Click OK and quit the operating software.

 Restart the operating software as required.