Operation Manual D-480

©BÜCH Labortechnik AG

10 - Menu and dialog descriptions

312

If you are using multiple analyzers with one PC, you must specify one instrument identifier on the
"Identifier" tab so that the LIMS data can be allocated to the correct instrument. You can also
define user names.

Note

Note that the LIMS port name must not be identical to the port name between the PC and the
analyzer. This could lead to a system crash.

Open

Open the dialog via Options > Settings > LIMS.

Elements

The following describes the elements of the dialog.

"LIMS data" tab

In section "Data to send":

Label

Element type

Meaning

{Data member}

Checkbox

Sends the corresponding data member.

Send data
automatically after
every analysis

Checkbox

Automatically send the data to LIMS after every analysis.

Send weighing data
on input/transfer

Checkbox

After entering or transferring the sample weight, it will be
sent automatically to the LIMS.

Note: Sends every element with a check mark. The order corresponds to the list from top to bottom. If
you change the selection of elements to send, you must edit the LIMS software, otherwise the data
may be misinterpreted.

"Communication" tab:

In section "Data transfer":

Label

Element type

Meaning

none

Radiobutton

No data are sent to LIMS.

via serial port

Radiobutton

Data are transferred via the serial port.

via network

Radiobutton

The data to send to the LIMS are written to a file.

The file can be accessed from the LIMS and from the
instrument operating software.

In section "Serial":

Label

Element type

Meaning

Port name

Textbox

Field to input the name of the interface.

The port name must not be the same as the port for
communication between the PC and analyzer. Otherwise,
operation of the analysis system will be disrupted.

Baud rate

Textbox

Field to input transfer parameters.

Databits

Textbox

Field to input transfer parameters.

Stopbits

Textbox

Field to input transfer parameters.

Parity

Textbox

Field to input transfer parameters.

Note: The textboxes in the "Serial" section only have to be filled in if you have selected the "via serial
port" radio button in the "Data transfer" section. The settings here must match those of the LIMS
interface.