Removing the quartz glass bridge, Cleaning and installing the quartz glass bridge, O_2742 – BUCHI DuMaster D-480 User Manual

Page 214: O_2803, O_3742

Advertising
background image

Operation Manual D-480

©BÜCH Labortechnik AG

8 - Maintaining the instrument

214

Removing the quartz glass bridge

Process

The section is divided into the following phases:

Preparing the instrument, gas pressure

Removing the quartz glass bridge

Furnace area

The following picture shows the furnace area in detail:

Legend:

1 Quartz glass bridge

Phase 1, Preparing the instrument, gas pressure

Safety instruction

Strictly observe the safety instruction "Gas pressure" under Warning notes during operation (on
page 24).

To prepare the instrument for maintenance work follow the instructions in Preparing and following
up maintenance work
(on page 171).

Phase 2: Removing the quartz glass bridge

Proceed as follows:

 Open the front door.

 Remove the heat cover on the quartz glass bridge under the furnace.

 Remove the ground-in clamps from both the combustion and post combustion tube (bottom side)

while holding the quartz glass bridge.

 Detach the quartz glass bridge and place it on a non-combustible surface.

 Protect the hot quartz glass bridge from unauthorized access.

Cleaning and installing the quartz glass bridge

What contaminations are possible?

Black residue (tar) indicate poor combustion and incorrect O

2

dosing.

Advertising