10 - Menu and dialog descriptions

316

Label

Element type

Meaning

O2 dosing time

Textbox

Field to input the duration of O

2

dosing during the

combustion phase.

O2 dosing flow

Textbox

Field to input the duration of oxygen flow during the
combustion phase.

O2 cut off threshold

Textbox

Field to input a value which delimits the oxygen excess. If
during the oxygen dosing the measured oxygen
concentration exceeds this value, then the oxygen supply
will be turned off.

Autozero delay

Textbox

Field to input a time. Defines the time to wait after
introduction until the system performs autozero alignment.

Peak anticipation

Textbox

Field to input a time. Defines the time the system waits for
the occurrence of the nitrogen peak. If the system has not
detected a peak start after expiration of this peak
expectation time, it assumes that the sample does not
contain nitrogen.

Integrator reset delay Textbox

Field to input a time. Defines the time after which the
integrator is reset to zero before beginning actual
integration.

Kjeldahl factor

Textbox

Field to input a factor which is used to normalize the
nitrogen results measured by the rapid N to their Kjeldahl
equivalent values.

New

Button

Allows the definition of a new method. This function
empties the input boxes, thus data for a new method can
be entered.

Delete

Button

Deletes a method previously selected in the listbox.

Default method

Button

Defines the method selected in the list box as the default
method. If during input of sample data no method has been
specified then the program suggests the name of the
default method.

Save/

Create

Button

Saves changes of an existing method or creates and saves
the settings of a newly defined method.

Close

Button

Closes the dialog.

Further information

Further information includes instructions in which the dialog is used and knowledge articles
concerning the dialog:

Viewing method settings (on page 131).

Defining custom methods (on page 131).

Copy methods (on page 133).

Optimizing basic instrument settings (on page 99).

Error handling (Dialog)

Function

In this dialog you can define how you want the operating software to react to detected errors.

Open

Open the dialog via Options > Settings > Error handling.