Configure calibration (dialog), H_860, O_4122 – BUCHI DuMaster D-480 User Manual

Page 336

Advertising
background image

Operation Manual D-480

©BÜCH Labortechnik AG

10 - Menu and dialog descriptions

336

Label

Element type

Meaning

Cancel

Button

Discards calibration coefficients that have just been
computed or input.

Further information

Further information includes instructions in which the dialog is used and knowledge articles
concerning the dialog:

Viewing calibration coefficients (on page 177).

Defining standard substances as calibration samples (on page 122).

Performing calibration.

Background knowledge required for calibration (on page 52).

Configure calibration (Dialog)

Function

In this dialog you can specify the calibration curve calculation methods. It is not possible to perform
the calibration comuptation without this information.

Open

Open the dialog via Calibration > Calibrate. In order to open the dialog, sample memory must
contain a sufficient number of valid calibration samples.

Elements

The following describes the elements of the dialog.

On the respective tab:

Label

Element type

Meaning

One calibration range Radiobutton

Only one calibration curve is computed over the entire
measuring range for the respective element.

Polynomial degree

Textbox

Field to input the polynomial degree for the calibration
curve calculation over the entire measuring range. May be
set from 1 to 4.

Two calibration
ranges

Radiobutton

For each element, the whole measuring range is divided
into an upper and lower calibration range. A separate
calibration curve is computed for each calibration range.

Polynomial degree,
lower range

Textbox

Field to input the polynomial degree for the calibration
curve calculation in the lower range. May be set from 1 to
4.

Polynomial degree,
upper range

Textbox

Field to input the polynomial degree for the calibration
curve calculation in the upper range. May be set from 1 to
4.

Remaining elements:

Label

Element type

Meaning

OK

Button

Continues with the calibration calculation. If the option
"Two calibration ranges" has been checked for at least for
one element, then the "Specify lower and upper calibration
range" dialog will open. Here, the split-up of both
calibration ranges can be carried out.

Cancel

Button

Aborts the process.

Advertising