Performing measurements, O_3673, O_3674 – BUCHI DuMaster D-480 User Manual

Page 138

Advertising
background image

Operation Manual D-480

©BÜCH Labortechnik AG

7 - Using the instrument

138

Animal feed, approx. 500 mg medium fat content 50 mgFeed

250 ml/min, 100 sec

Cheese, sausage, appr. 500 mg high fat content 500fatCheeseSausage

300 ml/min, 150 sec

Blank value with O

2

Blank with O2

80 ml/min, 60 sec

Blank value without O

2

Blank without O2

50 ml/min, 1 sec

Beer, milk, soil, appr. 500 mg medium fat content 500mgBeerMilkSoil

120 ml/min, 60 sec

Starch, approx. 400 mg

400mgStarch

250 ml/min, 150 sec

Soil, approx. 250 mg

250mgSoil

100 ml/min, 60 sec

Optimize O

2

dosing

If your sample matrix is always the same, it is advisable to optimize oxygen dosing for reasons of
costs. Proceed as follows:

 Measure a typical sample of known element content several times with the same sample weight

(max. ± 2 mg fluctuation).

 Systematically increase O

2

cut off threshold until you get a false low reading or a substantial

widening of the N peak.

 Set the O

2

cut off threshold of the last successful measurement before the false low reading

occurred. These settings are optimal for this matrix and sample weight.

 Check whether oxygen dosing is generally usable. To do so, measure several known samples of

similar matrix.

Example for optimization of oxygen dosing

O

2

cut off threshold = 30 %, total O

2

amount = 589 ml

O

2

cut off threshold = 40 %, total O

2

amount = 478 ml

Performing measurements

What you need to know

You must have read the Oxygen dosing values (on page 137) section so that you can set the
correct values for measuring.