Operation Manual D-480

©BÜCH Labortechnik AG

10 - Menu and dialog descriptions

321

Safety instruction

Strictly observe the safety instruction "Improper changes" under Warning notes during operation
(on page 24).

Note

This section only describes the sections of the tabs in which you can make changes via the "Device
configuration" dialog only, as these sections do not have their own settings menu.

If 21 CFR Part 11 functionality is enabled, every change to the configuration must be justified in the
"Comment modification" dialog.

Open

Open the dialog by selecting Options > Configuration.

Elements

The following describes the elements of the dialog.

Valid for all tabs:

Label

Element type

Meaning

Section

Dropdown listbox This field lists the available sections. Field to input a name for

a new section.

Entries

Listbox

This field lists the entries of a previously selected section.

New Section

Button

Creates a new section under the name previously entered in
the dropdown listbox.

Edit

Button

Opens the "Configure key value" dialog. Allows you to edit a
section entry previously selected from the list.

Delete

Button

Deletes a section previously selected in the dropdown listbox.
Deletes a section entry previously selected in the listbox.

Paste

Button

Opens the "Configure key value" dialog. Allows you to insert
new entries in a section.

Import

Button

Opens the "Open" dialog. Allows you to import further
sections.

Export

Button

Opens the "File save as" dialog. Allows you to export
sections.

Delete all

Button

Deletes all entries which are included in the activated tab.

OK

Button

Saves new settings

"DEVICE.INI" tab

The following table describes the entries under the "AnalyzerPort" section:

Entry

Standard

Description

Name

COM1

Defines the serial interface between the PC and the analyzer.

Baud

9600

Defines the transfer rate of the serial interface between the PC
and the analyzer.

Data

8

Defines the number of databits of the serial interface between the
PC and the analyzer.

Stop

1

Defines the number of stopbits of the serial interface between the
PC and the analyzer.