O_3751, O_2560 – BUCHI DuMaster D-480 User Manual

Page 231

Advertising
background image

Operation Manual D-480

©BÜCH Labortechnik AG

8 - Maintaining the instrument

231

Note: When you close the ground-in ball-and-socket joints, you must be able to feel an elastic
resistance when you press the o-rings into the pans, otherwise you must replace the o-rings.

 Place the ungreased seal on the combustion tube connection (3) at the top.

 Check whether the seals are in the pans of the plugs of combustion and post combustion tube.

Furnace area

The following picture shows the details of the furnace area:

Legend:

1 Handle
2 Bayonet catch
3 Reduction tube connection
4 Post combustion tube connection

Phase 1: Installing the reaction tubes, continued

Proceed as follows:

 Push the furnace back:

Lift the bayonet catch (2) slightly so as not to damage the seal on the combustion tube
connection.

Close the bayonet catch by turning the bayonet catch handle (1) to the left.

 Leave the end of the reduction tube open.