8 - Maintaining the instrument

181

 If deleting has not had the desired effect, you can undo it. To do so, select the deleted sample

and click the "Include/Exclude"

button again. The sample is now included again.

 Once you have excluded all samples with significant deviations from the calibration curve, click

the "Next"

button. The next calibration curve appears in the calibration view.

 Once you have optimized the last calibration curve and clicked the "Next"

button, the

"Calibration coefficients" dialog is displayed.

 Check...

... the calibration coefficients "a" to "e".
Note:
For the lower calibration range it is important that coefficient "a" is 1x10

-3

... 3x10

-2

or less.

Coefficient "a" corresponds to a calibrated blank value and should be as small as possible.

... the possible working range defined in "Min." and "Max.".

 Decide:

If...

... you accept the calibration coefficients: Click OK. The calibration coefficients are applied.

... you do not accept the calibration coefficients: Click Cancel. Calibration is aborted.