11. Technical information

General points

P

ower supply NIMH rechargeable

battery (NIMH) (4,8 V ª 1200 mA/h)

C

hargers: The only chargers used for

recharging the battery bear the follo-
wing information:

Type TR503-02-E-133A03
Input 90-264 VAC / 47-63Hz / 0.5A max.
Output 9V / 400mA / 6W

Type TR1509-06-E-133A03
Input 90-264 VAC / 47-63Hz / 0.5A max.
Output 9V / 1.4A / 15W

Type TR503-02-A-133A03
Input 90-264 VAC /47-63Hz / 0.5A max.
Output 9V / 400mA / 6W

Type TR1509-06-A-133A03
Input 90-264 VAC / 47-63Hz / 0.5A max.
Output 9V / 1.4A / 15W

Type TR503-02-U-133A03
Input 90-264 VAC / 47-63Hz / 0.5A max.
Output 9V / 400mA / 6W

Type TR1509-06-U-133A03
Input 90-264 VAC / 47-63Hz / 0.5A max.
Output 9V / 1.4A / 15W

Neurostimulation

A

ll electrical specifications are given

for a charge between 500 and 1000
ohms per channel.

Channels: four independent and individually
adjustable channels electrically insulated from
each other and from ground

Impulse shape: constant rectangular current
compensated to eliminate any direct current
component to avoid any residual skin polarisation

Maximum current of an impulse: 100 milliamperes

Intensity increase interval:
– Manual stimulation intensity adjustment: 0-
694 (energy units)
– Impulse duration: 0.5 mA
From 50 to 400 microseconds

Maximum quantity of electricity per impulse: 80
microcoulombs (2 x 40 µC compensated)

Typical rise time for an impulse
3 microseconds (between 20% and 80% of
maximum current)

Impulse frequency
from 1 to 150 Hertz

8. Standards

T

he Compex is based directly on me-

dical technology.

T

o guarantee your safety, the design,

manufacturing and distribution of
Compex are in conformity with the
requirements of the European Direc-
tive 93/42/CEE.

T

he device is in conformity with the

standard for general safety rules for
electromedical devices IEC 60601-1,
the standard for electromagnetic
compatibility IEC 60601-1-2 and the
standard for special safety rules for
nerve and muscle stimulators IEC
60601-2-10.

C

urrent international standards (IEC

60601-2-10 AM1 2001) require that a
warning be given concerning the ap-
plication of electrodes to the thorax
(increased risk of cardiac fibrillation).

D

irective 2002/96/CEE Waste electri-

cal and electronic equipment
(WEEE).

9. Patent

S

nap electrode: patent pending.

10. Normalised symbols

I

mportant: Under some condi-

tions, the effective figure for the

stimulation pulses can exceed 10 mA
and 10 V. The information given in this
handbook must be strictly observed.

T

he Compex is a class II device

unit with its own internal electric

power, with type BF applied sections.

W

aste electrical and electronic

equipment (WEEE).

T

he “On/Off” switch is a multi-

function key:

Functions

Symbol N°(according to CEI 878)

On/Off (2 positions, stable)

01-03

Waiting or preparation

for part of the unit

01-06

Stop (switching off)

01-10

EUROPE