Description

callout

Access Level control set to Operator. By default, this is the
level required to assign exclusive control to the handset.

1

Exclusive Control selector for the Handset interface enabled.
When the handset has exclusive control, an asterisk (*) is

2

displayed before the display page code. In this example, the
display page is Manual Command, shown as “*MC.”

Exclusive Control selector for the Station Manager
application interface is disabled when the handset is
assigned exclusive control.

3

It is also disabled when you open Station Manager
application, as neither the handset nor the PC interface is
assigned exclusive control by default.

This is important because it prevents anyone from changing tuning and calibration settings at the PC while
the operator is using the handset to move the actuator. While you do not use tuning and calibration controls
to move the actuator directly, they do affect actuator response.