MTS Series 793 User Manual

Page 221

Advertising
background image

Note:

Transducer ID modules are not available for FlexTest IIm/CTC/CTM, FlexTest 40, FlexTest 60, FlexTest
100, or FlexTest 200 controllers.

A transducer ID module includes:

A transducer ID circuit with calibration information

A shunt-calibration resistor

Up to three bridge-completion resistors

How to Assign a Transducer ID Module to a Sensor Input

Transducer ID modules can only be used with Series 493 Conditioners.

1. Connect a transducer ID module to the sensor.

2. Display the Station Setup window.

a) On the Station Manager window toolbar, select an access level of Tuning.

b) On the Station Manager window Display menu, click Station Setup.

3. In the Station Setup window navigation pane, select the sensor input where the Transducer ID module is

connected.

4. Change the sensor input sensor-file assignment to <Transducer ID>.

a) In the Station Setup window navigation pane, select the sensor input where the Transducer ID module

is connected.

b) Click the Sensor tab.

c) In the Sensor File list, click <Transducer ID>.

d) Click Show Trans ID to display the Transducer ID contents. Verify that the sensor assignment is correct.

e) Click Assign.

MTS Series 793 Control Software 221

Station Setup

Advertising