How to shift a conditioner’s electrical zero – MTS Series 793 User Manual

Page 212

Advertising
background image

Description

Callout

After installing our specimen, the feedback is saturated for Range 2,
the smaller range.

1

You can shift the electrical zero to bring the Range 2 feedback out of
saturation.

2

Range 2 (±1 cm)

3

Sensor Feedback = + 1.3 cm saturated feedback for Range 2 (±1 cm)

4

Shift the electrical zero

5

Sensor Feedback = 0 cm at installed position after shifting zero

6

How to Shift a Conditioner’s Electrical Zero

Before shifting the conditioner’s electrical zero, you must first position the actuator at the desired location and
disable the station power or switch to another control mode.

212 MTS Series 793 Control Software