MTS Series 793 User Manual
Page 256

Item
Callout
Function Generator (or Drive File)
1
Desired
2
Compensation Filter
3
Drive
4
Test System
5
PID Controller
6
Actuator
7
LVDT
8
Position Feedback
9
Inverse System Identifier
10
Response
11
AIC Tab Settings
Description
Item
Sets how quickly compensation values are applied to converge the feedback
and command signals.
Convergence Rate
Sets a time delay between the desired and response signals.
Anticipation
Sets the length of the adaptive filter, which influences how completely the
dynamics of the controller/actuator/specimen combination are cancelled.
Impulse Response
Sets the crossover frequency.
Crossover Freq
When applying AIC to a dual compensation mode configured in the Station
Builder application for Amplitude Control Only, set this control to 0.
Allows the selection of a required type of pre-emphasis filter.
Pre-emphasis Filter
256 MTS Series 793 Control Software
Station Setup