Description

Item

Displays a list of the parameters and signals that are used in this expression.

Uses

Calculation Parameters Panel

This panel summarizes the current values used in calculated parameters.

Access

Display menu > Station Setup > Station Setup window > navigation pane > Calculations > Calculation
Parameters

The window’s access level setting determines a user’s ability to change these values.

Warning:

Small calculation parameter changes can produce large control-loop changes, resulting in violent
actuator movement.

Violent actuator movement can result in injury and equipment damage.

Use care when changing calculation parameters.

Calibration Tab for a Calculated Input (Channels)

This tab displays the expression used in the selected control mode that uses a calculated input.

Access

Station Setup window > navigation pane > Channels control mode using calculated input >

> Calibration

tab

Note:

Requires Calibration access level (or higher) to edit any item.

Calculation Tab for a Calculated Input

Description

Item

Specifies the selected inputs full-scale minimum and maximum values.
The result of this calculation gets clipped to these values.

Fullscale Min/Max

Displays the formula used to produce the input.

Expression

Use the Calculation Editor window to define the formula.

Displays a list of the parameters and signals that are used within this
expression.

Uses

384 MTS Series 793 Control Software

Calculated Signals