Sensor File Editor

How to Create a Sensor File

You can use the Sensor File Editor window to create a sensor file from any user access level. You must be
at the Calibration access level to edit an existing sensor file.

Note:

The Model 494.21 Elastomer DUC (AC mode) excitation frequency is set to a fixed value during
manufacturing and cannot be changed by software. When creating a sensor file for the Model 494.21
AC, make sure you enter the correct excitation frequency.

1. On the Station Manager application Tools menu, click Sensor File Editor.

2. On the Sensor File Editor toolbar, click Open, and then New.

3. In the Conditioner Type list, select the conditioner that is used with the sensor you are calibrating.

Description

Callout

Select the Conditioner Type.

1

Select the Dimension.

2

Define at least one range or turn on Full Scale conditioning, and
set its Fullscale Min/Max values.

3

4. Enter Basic TEDS information for the sensor that will use the sensor file (Series 494 Hardware only).