About electrical zero offset – MTS Series 793 User Manual

Page 208

Advertising
background image

Description

Item

Note:

If station power is on, Auto Zero and Manual Zero cannot be used
on active control modes.

Automatically applies an offset to zero the selected conditioner’s sensor
signal and zeros the manual offset settin

Auto Zero

Optional—Select the Electrical Zero Lock checkbox to disable Auto Zero
for the Tuning and Operator access levels.

Click Manual Zero to apply an offset to make the selected conditioner’s
sensor signal zero.

Manual Zero

Displays the current value (in engineering units) of the selected signal
including the offset.

Current Value

About Electrical Zero Offset

If you want to shift a small, calibrated sensor range away from the sensor’s calibrated electrical zero, you can
offset the sensor’s electrical output in the conditioner.

A traditional offset shifts the feedback signal and an electrical zero offset shifts the conditioner signal.

An electrical zero offset cannot be applied to a conditioner used in an active control mode with hydraulic
pressure applied.

Item

Callout

Sensor Voltage

1

Conditioner

2

Gain

3

Electrical Zero

4

Electrical zero is only available on certain types of conditioners.

Conditioner Voltage

5

Controller/Application

6

208 MTS Series 793 Control Software

Station Setup

Advertising