Description

Item

Click the Channel button to toggle between individual channels and a Master
Command Group.

Displays the control mode that is being adjusted.

Active Mode

Adjusts the program’s signal’s mean (setpoint).

Setpoint

When a Master Command Group is selected, this control changes the
setpoints of channels selected in the Channel Options window’s Master
Command tab.

Selecting a Master Command Group that has individual channels at different
Setpoint values changes this control. It displays the highest and lowest
Setpoint values for the Group. It also displays pointers that indicate the
highest and lowest Setpoint values. Pointers lock together once they have
been used to move the Group to a common Setpoint value.

Adjusts the program signal’s span.

Span

When a Master Command Group is selected, this control changes the spans
of channels selected in the Channel Options window’s Master Span tab.

Selecting a Master Command Group that has individual channels at different
Span values changes this control. It displays the highest and lowest Span
values for the Group. It also displays pointers that indicate the highest and
lowest Span values. Pointers lock together once they have been used to
move the Group to a common Span value.

Adjusts the span for all channels included in a master span group.

Master Span

This item is not displayed if Master Span is selected for display on the Station
Manager main window.

Note:

If necessary, click the Setpoint or Span text to display a window where you can change the slider-bar
ranges.

How to Adjust Setpoint and Span During a Test

The Setpoint controls can adjust the setpoint of both single channels and multiple channels.

Setpoint and Span controls in a Master Command Group are available only on a single channel.

Reducing the Station Controls panel’s Master Span setting further reduces an individual channel’s Span
setting. For example: a Master Span of 50% halves an individual channel’s Span of 50%, giving the channel
an effective 25% span.