How to set up aicx compensators – MTS Series 793 User Manual

Page 266

Advertising
background image

5. Set up a random noise program.

6. Find the initial impulse response.

7. Set Anticipation.

8. Train the forward model.

9. Compensate the system with the inverse model.

How to Set Up AICX Compensators

1. Select the Function Generator, and in the Compensator list, select AICX.

2. Access the compensator.

a) Select an access level of Tuning.

b) In the Station Setup window navigation pane, locate and select the Channels control mode that is to

receive AIC compensation.

c)

In the Station Setup window, click

d) On the Compensators panel, click the AICX tab.

Warning:

Changes in compensation values made with hydraulic pressure on can result in unexpected
actuator movement.

A moving actuator can injure anyone in its path.

Always clear the actuator area before changing compensation values.

3. Set initial AICX values.

a) Set Forward Convergence Rate, Inverse Convergence Rate, and Anticipation to zero.

b) Set Impulse Response to 0.5.

c) Keep Operating Band control settings at their default values. The default value is 0.00 - 100.00 Hz.

d) Set Adaptation State to Hold.

e) Reset the forward and inverse models by selecting Reset Fwd and Reset Inv.

4. Set up and enable limits to protect yourself and your equipment.

5. Install a dummy specimen.

6. Set up a random noise tuning program.

266 MTS Series 793 Control Software

Station Setup

Advertising