Readouts, About readouts, How to configure readouts – MTS Series 793 User Manual

Page 272

Advertising
background image

Warning:

Changes in compensation values made with hydraulic pressure on can result in unexpected
actuator movement.

A moving actuator can injure anyone in its path.

Always clear the actuator area before changing compensation values.

5. In the PVP tab, define PVP values:

a) Set the Convergence Rate.

b) Set the Sensitivity.

c) For Adaptation State, select Resume.

6. On the File menu, select Save Parameters to save values.

Readouts

About Readouts

Readout resources (typically, D to A converters) are used to send user-defined station signals to external
readout devices such as oscilloscopes and meters.

Readouts Summary Panel

This Station Setup window displays the real-time values of readout signals.

How to Configure Readouts

Use the Readouts Summary panel to select the signal that you want to monitor with external devices, such
as oscilloscopes and meters.

1. In the Station Setup window, select the readout that you want to configure.

2. Select a signal from the Signal list.

272 MTS Series 793 Control Software

Station Setup