Digital Input Actions

Digital Input Actions

Action

Setting

No action occurs. This setting dims the indicator.

Disabled

Writes a message to the Message Log and the indicator switches from white
to red.

Indicate

Stops any program command, writes a message to the Message Log, turns
off pressure at the hydraulic service manifold (HSM), and switches the indicator
from green to red.

Station Power Off

For single station systems: If the “Last Off” function is enabled in the .hwi file
or equivalent jumpering, the hydraulic power unit (HPU) is turned off. If the
“Last Off” function is disabled, the HPU will remain on.

For multistation systems: If the “Last Off” function is enabled in the .hwi file
or equivalent jumpering, the hydraulic power unit (HPU) will be turned off only
if the interlock occurs on the last station running. If the “Last Off” function is
disabled, the HPU will remain on.

Stops any program command, writes a message to the Message Log, clamps
the servovalve (if enabled in the .hwi file), turns off pressure at the hydraulic
service manifold (HSM), and switches the indicator from green to red.

Interlock

An Interlock must be reset to resume testing.

For single station systems: If the “Last Off” function is enabled in the .hwi file
or equivalent jumpering, the hydraulic power unit (HPU) is turned off. If the
“Last Off” function is disabled, the HPU will remain on.

For multistation systems: If the “Last Off” function is enabled in the .hwi file
or equivalent jumpering, the hydraulic power unit (HPU) will be turned off only
if the interlock occurs on the last station running. If the “Last Off” function is
disabled, the HPU will remain on.

Stops any program command, writes a message to the Message Log, and
forces a program interlock. Produces the same action as clicking the Program
Stop button on the Station Controls panel.

Program Stop Interlock

A Program Stop Interlock must be reset to resume testing.

Channel Low and Channel High trigger modes do not support this action.

Places a hold on any program command, writes a message to the Message
Log, and forces a program interlock. Produces the same action as clicking
the Program Hold button on the Station Controls panel.

Program Hold Interlock

A Program Hold Interlock must be reset to resume testing.

The Function Generator does not recognize Program Hold or Program Hold
Interlock, but instead goes to Stop.

Channel Low and Channel High trigger modes do not support this action.

278 MTS Series 793 Control Software