Description

Setting

Stops any program command, writes a message to the Message Log,
clamps the servovalve (if enabled in the .hwi file), turns off pressure

Interlock

at the hydraulic service manifold (HSM), and switches the indicator
from green to red.

Note:

An Interlock must be reset to resume testing.

For single station systems: If the “Last Off” function is enabled in the
.hwi file or equivalent jumpering, the hydraulic power unit (HPU) is
turned off. If the “Last Off” function is disabled, the HPU will remain
on.

For multistation systems: If the “Last Off” function is enabled in the
.hwi file or equivalent jumpering, the hydraulic power unit (HPU) will
be turned off only if the interlock occurs on the last station running. If
the “Last Off” function is disabled, the HPU will remain on.

Stops any program command, writes a message to the Message Log,
and switches the indicator from green to red. Produces the same action
as clicking the Program Stop button on the Station Controls panel.

Program Stop Interlock

Note:

A Program Stop Interlock must be reset to resume testing.

Places a hold on any program command, writes a message to the
Message Log, and switches the indicator from green to red. Produces

Program Hold Interlock

the same action as clicking the Program Hold button on the Station
Controls
panel.

Note:

A Program Hold Interlock must be reset to resume testing.

Note:

The Function Generator does not recognize Program Hold or
Program Hold Interlock, but instead goes to Stop.

How to Define the Excitation Failure Mode and Actions

If necessary, perform this procedure for each sensor input.

216 MTS Series 793 Control Software

Station Setup