MTS Series 793 User Manual
Page 298

Action
Setting
off only if the interlock occurs on the last station running. If the “Last Off”
function is disabled, the HPU will remain on.
Stops any program command, and writes a message to the Message Log
and switches the indicator from green to red. Produces the same action as
clicking the Program Stop button on the Station Controls panel.
Program Stop Interlock
Note:
A Program Stop Interlock must be reset to resume testing.
Places a hold on any program command and writes a message to the
Message Log and switches the indicator from green to red. Produces the
Program Hold Interlock
same action as clicking the Program Hold button on the Station Controls
panel.
Note:
A Program Hold Interlock must be reset to resume testing.
Note:
The Function Generator does not recognize Program Hold or Program
Hold Interlock, but instead goes to Stop.
298 MTS Series 793 Control Software
Station Setup