Advanced tab – MTS Series 793 User Manual

Page 239

Advertising
background image

Description

Item

Displays all signals available for clamping event. Move signals between the
Available to Included lists with the arrow buttons or by double-clicking signal
names.

Available/Included list

The digital I/Os in the Included list are used to determine the clamping state.
Any number of I/Os can be selected here.

Dropdown list displays all signals added to Included list.

Signal

Select High or Low for each signal selected above.

Trigger

Advanced Tab

Access

Station Setup window > navigation pane > Channels PIDF/CLC control mode >

> Advanced tab

Advanced Tab

Description

Item

Sets the maximum integrator value as a percentage of full-scale output.

Integrator Limit

Sets the maximum controller output as a percentage of full-scale output
(up to a maximum of 500% full-scale output). This control is independent
of the Output Limit Enable control associated with the output resource).

Controller Output Limit

These settings determine the remote setpoint null pace tolerance band
(in absolute engineering units) for a displacement control mode. Each
displacement control mode has its own setpoint null pace band setting.

Remote Setpoint Null Pace Band

Remote setpoint null pace settings are saved in parameter sets.

When you use a setpoint device (such as a handset) to position the
actuator, any time the error exceeds the null pace band setting, the
setpoint command will stop. The setpoint will not resume until the feedback
catches up with the command and the error is within the setpoint null
pace band.

Remote setpoint null pacing does not affect Manual command, the
Setpoint slider control, or Park and Ride controls.

You must be at the Tuning access level to change this control.

Local Integrator Control

Station Select (default)—Sets the local integrator mode to the integrator
mode set with the Station Integrator Control (Tools > Station Options >
Station View Options). The station integrator is associated with all of the
channels (and associated control modes) in the station. The local
integrator is associated with the selected control mode.

The Auto-High and Auto-Low settings override the integrator mode set
with the Station Integrator Control.

MTS Series 793 Control Software 239