6.3.2 Offset Calibration to each Channel

@Description

To correct the offset errors of AD converter by using the offset
calibration.

@Syntax

$(Addr)1(Channel No)<CR>

$

Command leading code

(Addr)

Address ID

1 Offset

calibration.

(Channel No)

Channel for calibration.(1 character)0~2

@Response

!(Addr)<CR>

or

?(Addr)<CR>

!

Command is valid.

?

Command is invalid.

(Addr) Address

ID.

@Example

User command:

$0612<CR>

Response:

!06<CR>


To perform the offset calibration for analog input module
ND-6013 channel 2 at specified address 06 (Hex).

Note: To perform the calibration, a proper input signal should

be connected to the analog input module. Different
input range have different input voltage, detail refer
chapter 9 “Calibration” .