Calibration

9.1 How to Calibrate the Analog Input Modules ?

What do you need to do calibration ?
1.

One 5 1/2 digit multimeter.

2. A voltage calibrator or very stable and noise free DC

voltage generator.

3.

A precision resistance decade box or discrete resistors.

4.

NuDAM Administration Utility.

Calibration Procedure for ND-6017

1. Select the correct input range, different input range have

different apply calibration voltage.

2. Apply

the

correct

offset voltage to the analog input

module, detail voltage value, see table 9-2.

3. Send

Offset Calibration $(Addr)1” to analog input

module five times.

4.

Apply the correct span voltage to the analog input module,
detail voltage value, see table 9-2.

5. Send

Span Calibration $(Addr)0” to analog input module

five times.

6.

Repeat procedure 2 to procedure 5 two times.