Calibration Procedure for ND-6013 Firmware Rev A3.05

1. Select the correct input range, different input range have

different apply calibration resistance.

2. Apply the correct offset resistance to the analog input

module ND-6013 channel 0, detail resistance value, see
table 9-3.

3. Send “Offset Calibration $(Addr)10” to analog input

module ND-6013 channel 0 .

4. Apply the correct span resistance to the analog input

module ND-6013 channel 0, detail resistance value, see
table 9-3.

5. Send “Span Calibration $(Addr)00” to analog input

module ND-6013 channel 0 .

6. Apply the correct offset resistance to the analog input

module ND-6013 channel 1, detail resistance value, see
table 9-3.

7. Send “Offset Calibration $(Addr)11” to analog input

module ND-6013 channel 1.

8. Apply the correct span resistance to the analog input

module ND-6013 channel 1, detail resistance value, see
table 9-3.

9. Send “Span Calibration $(Addr)01” to analog input

module ND-6013 channel 1.

10. Apply the correct offset resistance to the analog input

module ND-6013 channel 2, detail resistance value, see
table 9-3.

11. Send “Offset Calibration $(Addr)12” to analog input

module ND-6013 channel 2.

12. Apply the correct span resistance to the analog input

module ND-6013 channel 2, detail resistance value, see
table 9-3.

13. Send “Span Calibration $(Addr)02” to analog input

module ND-6013 channel 2 five times.

14. Repeat procedure 2 to procedure 13 two times.