Calibration Procedure for ND-6013 Firmware Rev C4.60

1. Only open channel one and disable the other two channels.
2. Disable open detect function.
3. Select the correct input range, different input range have

different apply calibration resistances.

4. Apply the correct offset resistance to channel 0, detail

resistance value, see table 9-3.

5. Send “Offset Calibration $(Addr)1” to analog input

module five times.

6. Apply the correct span resistance to channel 0, detail

resistance value, see table 9-3.

7. Send

Span Calibration $(Addr)0” to analog input module

five times.

Repeat procedure 3 to procedure 7 two times.

Calibration Procedure for ND-6018 Firmware Rev B1.10

1. Disable all the channel and open detect function.
2. Select the correct input range, different input range have

different apply calibration voltage.

3. Apply the correct offset voltage to channel 0, detail

voltage value, see table 9-1.

4. Send

Offset Calibration $(Addr)1” to analog input

module five times.

5. Apply the correct span voltage to channel 0, detail

voltage value, see table 9-1.

6. Send

Span Calibration $(Addr)0” to analog input module

five times.

Repeat procedure 3 to procedure 6 two times.

Calibration Procedure for ND-6018 Firmware Rev E1.00

Power off the module.
1. Connect the default pin to GND.(Because the calibration

procedures must be run under default mode).

2. Power on the module and use utility searching the module.
3. There must a “*” before the address ID.
4. Disable open detect function by sending command

$00O0”.

5. Only enable Ch0.