ADLINK ND-6067 User Manual

Page 271

Advertising
background image

269

Code

Input Range

Span

Calibration

Resistance

Offset

Calibration

Resistance

20

Pt-100, -100

°C to

+100

°C, α=.00385

200

Ω 50Ω

21

Pt-100, 0

°C to

+100

°C, α=.00385

200

Ω 50Ω

22

Pt-100, 0

°C to

+200

°C, α=.00385

200

Ω 50Ω

23

Pt-100, 0

°C to

+600

°C, α=.00385

350

Ω 50Ω

24

Pt-100, -100

°C to

+100

°C, α=.003916

200

Ω 50Ω

25

Pt-100, 0

°C to

+100

°C, α=.003916

200

Ω 50Ω

26

Pt-100, 0

°C to

+200

°C, α=.003916

200

Ω 50Ω

27

Pt-100, 0

°C to

+600

°C, α=.003916

350

Ω 50Ω

28

Ni-100, 0

°C to

+100

°C

200

Ω 50Ω

29

Ni-120, 0

°C to

+100

°C

350

Ω 50Ω

2A

0 Ohm to 60Ohm

60

Ω

0

Ω

Table 9-3: ND-6013 Calibration Resistance