6.3.3 Span Calibration to each Channel

@Description

To correct the gain errors of AD converter by using the span
calibration.

@Syntax

$(Addr)0(Channel No)<CR>

$

Command leading code (1 character)

(Addr)

Address ID (2 character)

0

Span calibration (1 character)

(Channel No)

Channel for Calibration (1 character) 0~2

@Response

!(Addr)<CR>

or

?(Addr)<CR>

!

Command is valid.

?

Command is invalid.

(Addr) Address

ID.

@Example

User command:

$0601<CR>

Response:

!06<CR>


To perform the span calibration for analog input module
ND-6013 channel 1, address ID is 06H.

Note:

To perform the calibration, a proper input signal should
be connected to the analog input module. Different
input range have different input voltage, detail refer
chapter 9 “Calibration” .