Sash, Operation, Adjustment – Labconco PURICARE Procedure Station Models 34812xx User Manual

Page 104

Advertising
background image

Page 104 of 197

Sash

Operation

The Logic sash uses a deadweight-counterbalanced system. The system consists
of two separate weights on either side of the liner. Both weights are
synchronized by a locked axle-pulley system. Both pulleys are fixed to the axle so
that both weights must move together, preventing the sash from racking.

Adjustment


There are no adjustments for the width of the sash track. There should be
approximately 1/8-inch lateral play of the sash in its track. If there is no lateral
play, examine one or both of the corner posts to ensure that they are not bent
inward. If the sash becomes racked, or is closing unevenly, it can be aligned as
follows:

1. Close the sash completely, until the lower edge of the sash is contacting

the sash stop. If the other end of the sash is not resting on its sash stop,
CAREFULLY lift the counterweight on the “high” side of the sash. This will
allow the high side of the sash to lower to the sash stop, leveling the
sash. See Figure 7-5.

Figure 7-5



Using pliers to lift a sash
weight to level the sash.


CAUTION: Exercise
caution when lifting a
counterweight. There is
a pinch potential
between the sash cable
and pulley.



Advertising