UV Lamp Timer

The UV Lamp Timer allows you to decide whether the UV lamp will operate continuously
or for a time period that you select.

Interval

When the “Interval” option is selected, the UV Interval Timer menu is displayed.

UV Interval Timer

This window lets you set the amount of time the UV lamp to stays lit
when activated. Use the “▲” and “▼” buttons to select the amount of
time you need, and then press the “OK” button to set the interval.

NOTE: Raising the sash will reset the timer to its original setting.
Pressing “UV Light” will turn the light off, pressing
“Mute/OK“will pause the timer.

Continuous

In the continuous mode, the UV light will remain lit until the “UV Light” button is
pressed, or the sash is raised.

UV Lamp Maintenance

This selection allows you to check how many hours the UV lamp has burned. It also
allows you to set the number of hours you want the lamp to last, and to reset the UV
lamp hourmeter.

UV Lamp Hourmeter

This display only shows how many hours the UV lamp has been lit, and how many
hours remain until you will receive a warning to replace the lamp.

Change UV Lamp Life

In this option, you can set the number of operating hours before receiving the
replace UV lamp warning. For most UV lamps, the output of UV light decreases at a
constant rate. Typically, after 6,000 hours of operation the lamp will output 80% of
the UV light it did when it was new. This option allows you to set the warning at a
life you prefer, in 100 hour increments.

Reset UV Lamp Clock

This option lets you reset the UV hourmeter to 0 whenever the lamp has been
replaced.