9. Route the sensor cable along the front of the cabinet and secure it to the

existing harness running along the front of the cabinet, using nylon ties along
the top of the cabinet as shown. Ensure that there is slack in the sensor cable
where it connects to the harness.

NOTE: Ensure the sensor harness does not contact any
moving parts.
Existing Wiring Harness

One nylon tie
(more ties are shown below)
Slack in the sensor cable


10. Remove the two Phillips screws on the front edge of the right side corner post,

and allow the front of the side panel to swing out and lift off.

11. Continue routing the sensor cable

along the main harness on the right
side of the cabinet. Secure the
sensor harness with the remaining
ties as shown.