Figure 5-16a & b




The Sensor Distance, as described in Table 5-4,
using the slots cut in the 6-foot template.

4.00-inch

mark

3.25-inch

mark




The Sensor Distance, as described in Table 5-4,
using a scale.



Table 5-4

Secondary Inflow Test Specifications

Model

34308 34300 34408 34400 34508 34500 34608 34600

Restricted Sash Opening

(inches)

4

4 4 3 3

2.25

2.25

4.25

4.25

Sensor distance (inches)

5

4 4 3.25 3.25 3.25 3.25 4 4

# of Test points

6

6

8

8

10

10

12

12

Test point location

*

*

*

*

*

*

*

*

Nominal Avg. Inflow Vel.

(FPM)

213 266 283 354 353 441 231 289

Avg. Inflow Vel. Range

(FPM)

203-223 253-278 269-296 337-371 336-370 420-462 220-242 275-302

Correction Factor (CF)

1

1

1

1

0.96

1

1.83

1.83

Average Inflow Volume

Range (AIV) (Avg. velocity

x CF)

203-223 253-278 269-296 337-371 336-370 420-462 403-443 503-553

Sash Open Area (Sq. Ft)

2.03 2.53 2.69 3.37 3.36 4.2 4.03 5.03

Inflow Velocity Range
(AIV / Sash open area)

100-110 100-110 100-110

100-110 100-110

100-110 100-110

100-110

1.

Locate the single row of holes at the front of the grille. Mark the 6

th

hole from

the side wall and subsequent test points every 9 holes until the number of test
points marked equals the width of the cabinet in feet (for a 3-foot cabinet, for
example, mark the first 3 points). Repeat for the opposite side, as shown in
Figure 5-17.

2.

Start the Purifier, and let it operate for at least 5 minutes. Establish the
necessary correction factor to the thermal anemometer to ensure compliance
with its performance in a calibrated wind tunnel. This data should be available
from the calibrator of the thermal anemometer.