Labconco PURICARE Procedure Station Models 34812xx User Manual

Page 166

Advertising
background image

Page 166 of 197

7. Install the new control board by reversing the above steps. Before closing and

securing the side panel, ensure the control circuit board display is centered on
the display window on the cabinet’s side wall. If the display is not centered,
loosen both mounting nuts, and move the control circuit board and its metal
frame until it is centered, then retighten the nuts.


8. Install the sensor assembly into the rear of the canopy connection as shown. Use

the supplied lockwashers and acorn nuts to secure the sensor assembly to the
canopy.



Canopy Air Sensor installation, Side View

Secure the sensor assembly with two lockwashers
and acorn nuts.

NOTE the orientation of the sensor assembly in the
canopy







Canopy Air Sensor installation, Side View

On 6-Foot Cabinets ONLY, ensure the
Sensor cable is secured to the top canopy, using
the wiring harness loops and screws provided.

NOTE: On 6-foot units, ensure the sensor cable
routing does not block or obstruct the operation
of the canopy relief valve on this side of the
canopy.

Advertising