Labconco PURICARE Procedure Station Models 34812xx User Manual

Page 72

Advertising
background image

Page 72 of 197

8. Remove the diffuser by raising the sash as high as it will go, and then removing

the two acorn nuts with a 7/16” wrench or socket, as shown in Figure 5-9.

Figure 5-9


Removing the diffuser acorn nuts (Dress
panel and lamps removed for clarity
only).










Note: The diffuser acorn nuts are secured to the studs with a removable thread locking
compound to prevent them from vibrating loose during shipment of the Logic. There
may be some resistance the first time the nuts are removed; this is normal. No
additional thread locking compound need be applied to these acorn nuts unless the unit
is being prepared for shipment.

9. Lower the front edge of the diffuser until clears the bottom of the sash, and then

pull the diffuser straight out of the cabinet, as shown in Figure 5-10.

Figure 5-10
















10. Open the following number of Laskin nozzle(s):

For all 3-foot Logic models – 1 nozzle

For all 4-, 5- and 6-foot Logic models – 2 nozzles

11. Ensure the pressure gauge on the generator is reading 20 +/- 1 PSIG.