Quickchart - logic type b2 biohazard cabinets – Labconco PURICARE Procedure Station Models 34812xx User Manual

Page 20

Advertising
background image

Page 20 of 197

QuickChart - Logic Type B2 Biohazard Cabinets

Model 34410

34610

Type

B2

B2

Cabinet Size (in feet)

4

6

Sash Opening (inches)

8

8

Starting Serial #

0705_

0705_

Downflow Data

Nominal Avg. Downflow (FPM)

55+/-5

55+/-5

Downflow test grid # of points

24

36

Test Grid (# rows x # columns)

3 x 8

3 x 12

Test Grid Data (in inches)

Grid distance from back & sides

6.0

6.0

Distance between rows

5.75

5.75

Distance between columns

5.21

5.50

Inflow Data

Nominal Average Inflow (FPM)

105+/-5

105+/-5

Sash Open Area (Sq. Ft)

2.69

4.03

Nominal Avg. Inflow Vol. (CFM)

283 +/-13

423+/-20

Avg. Inflow Vol. Range (CFM)

269-296

403-443

Secondary Inflow Data

3

Restricted Sash Opening (inches)

4

3

4.25

Sensor distance (inches)

5

4

4

# of Test points

8

12

Test point location

*

*

Nominal Avg. Inflow Vel. (FPM)

283

212

Avg. Inflow Vel. Range (FPM)

269-296

202-222

Correction Factor (CF)

1

2

Average Inflow Volume Range (AIV) (Avg. velocity x CF)

269-296

403-443

Sash Open Area (Sq. Ft)

2.69

4.03

Inflow Velocity Range (AIV / Sash open area)

100-110

100-110

Exhaust Data

DIM Exhaust Volume (CFM)

6

665

998

Traverse Exhaust Volume (CFM)

7

810

1218

Differential Pressure (in. H

2

0)

8

1.4

1.4

Exhaust (Inflow) Alarm Data

Alarm Setpoint (CFM Inflow)

9

150 +/- 5

223 +/- 5

Supply HEPA Filter Leak Test Data

Air Displacement (CFM)

445

665

Laskin Nozzles needed

1

1

Theoretical aerosol conc. (ug/l)

1

30

20

Actual aerosol conc. (ug/l)

2

15

12

Nominal Speed Control Setting

70

70

Supply HEPA Data

Width (inches)

48

72

Depth (inches)

18

18

Height (inches)

10

3.06

3.06

Performance (CFM)

445

665

Performance (Pressure in. H

2

O) .43-.57

.43-.57

Labconco P/N

3838401

3838403

Filtration Group P/N

42546L

42666L

AAF P/N

907-543-

202

907-543-

204