Purifier type b2 logic service operations, Recommended tools for service, External dress components – Labconco PURICARE Procedure Station Models 34812xx User Manual

Page 121: Front dress panel removal and installation

Advertising
background image

Page 121 of 197

Section Eight –

Purifier Type B2 Logic Service Operations


This section will review the tools needed, and common service operations for the Logic
Total Exhaust.

Recommended Tools for Service


The tools needed for most common service operations are:

9

Logic Certifier Kit, Labconco Part # 3858400

9

#2 Phillips Screwdriver or long bit for cordless drill

9

½-inch Socket with extension and ratchet

9

½-inch nut driver

9

7/16-inch socket or nut driver

9

5/16-inch socket or nut driver


Optional tools that may be needed include:

9

Cordless screwdriver/drill

9

Cordless ratchet

9

3/8-inch nut driver or socket and ratchet

9

Pliers

9

10-24 Tap

9

5/16 x 18 Tap


Note: If you use a cordless drill with an adjustable torque setting, set the
torque as low as possible to properly drive the screw or bolt.

External Dress Components

Front Dress Panel Removal and Installation

Figure 8-1

1.

Locate and remove the two Phillips screws that
secure the front panel as shown in Figure 8-1.
They are located on the bottom corners of the
front

dress

panel.

2.

Swing the bottom of the dress panel out to
clear the fluorescent light and then lift the
front dress panel straight up and away from
the

cabinet.

Advertising