Labconco PURICARE Procedure Station Models 34812xx User Manual

Page 177

Advertising
background image

Page 177 of 197

On Models Manufactured After April 2009 –


1. Turn off the Biosafety Cabinet.

2. Remove the exhaust HEPA filter cover by loosening the four Phillips

Screws on the sides of the cover, then pivot the front of the cover up, and

then pull it forward as shown.



Pivot the front of the cover up and then pull it

forward (dress panel removed for clarity only).


Loosen these screws only
Note: exercise caution when removing or installing the exhaust filter cover, to

prevent damaging the Exhaust HEPA filter.


3. Locate and remove the four acorn nuts

and lockwashers that secure each exhaust

cover bracket to the top of the Logic.

Keep the nuts and washers; you may
discard the brackets.



Advertising