On Models Manufactured After April 2009 –
1. Turn off the Biosafety Cabinet.
2. Remove the exhaust HEPA filter cover by loosening the four Phillips
Screws on the sides of the cover, then pivot the front of the cover up, and
then pull it forward as shown.
Pivot the front of the cover up and then pull it
forward (dress panel removed for clarity only).
Loosen these screws only
Note: exercise caution when removing or installing the exhaust filter cover, to
prevent damaging the Exhaust HEPA filter.
3. Locate and remove the four acorn nuts
and lockwashers that secure each exhaust
cover bracket to the top of the Logic.
Keep the nuts and washers; you may
discard the brackets.